With the resultant IEC standard 62494-1 “exposure index of digital X-ray imaging systems,” which provides a unified method to generate an exposure index value, requires users to input a target exposure index (EI_T). This offers the possibility to specify the noise level that is diagnostically appropriate for an examination and indicates a deviation index (DI) value that gives feedback to the user regarding technique and image quality based on signal-to-noise ratio. These are important values in providing feedback in regard to achieving optimal images at a low dose, the EI and DI don’t directly indicate patient dose.
 

“Expected value of the exposure index when exposing the image receptor properly“*

“May depend on the type of detector, the type of examination, on the diagnostic question…“*

*Source: IEC 62492-1 (International Electrotechnical Commission)

You are about to visit a Philips global content page

Continue

Our site can best be viewed with the latest version of Microsoft Edge, Google Chrome or Firefox.

You are entering a Philips Healthcare Australia website
Please select the checkbox

The information on this site is not intended for consumers. The information is directed exclusively to health professionals, health practitioners, persons who are purchasing officers in hospitals, and persons who are engaged in the business of wholesaling therapeutic goods (as per s42AA of the Therapeutic Goods Act 1989 (Cth), and s6 of the Therapeutic Goods (Therapeutic Goods Advertising Code) Instrument 2021).